Health product recall

BIOFIRE® Joint Infection Panel – 30 Pack

Last updated

Summary

Product
BIOFIRE® Joint Infection Panel – 30 Pack
Issue
Medical devices - Performance
What to do

Contact bioMérieux Canada Inc. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

BIOFIRE® Joint Infection Panel – 30 Pack

0806526 (exp. 2027-03-24)

RFIT-ASY-0138

BIOFIRE® Joint Infection Panel – 30 Pack

0677326 (exp. 2027-03-13)

RFIT-ASY-0138

BIOFIRE® Joint Infection Panel – 30 Pack

1289226 (exp. 2027-05-05)

RFIT-ASY-0138

Issue

Beginning April 5, 2026, elevated rates of KPC unexpected positives were observed in reagent final QC testing of the BIOFIRE® FILMARRAY® Pneumonia Panel, BIOFIRE® FILMARRAY® Pneumonia Panel Plus, and BIOFIRE® Joint Infection Panel, including in passing product that met final QC negativity requirements. Internal testing of failing product has reproduced KPC unexpected positive results, suggesting KPC nucleic acid (organism or synthetic) contamination in the pouch at levels that can lead to false positive results. Additional testing has confirmed that an isolated component, the outer primer pellet, is sufficient for producing unexpected positive KPC results.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies

Biofire Diagnostics, LLC.

515 Colorow Dr., Salt Lake City, Utah, United States, 84108

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82530