BIS Complete Monitoring System (2018-07-26)
- Starting date:
- July 26, 2018
- Posting date:
- August 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67386
Affected products
BIS Complete Monitoring System
Reason
Replacement instructions were revised for the Li-Ion battery packs used in BIS Vista Monitoring System. These revised instructions will require battery pack replacement after 4 years due to the potential for aged battery packs to experience an internal short that may result in a thermal event.
Affected products
BIS Complete Monitoring System
Lot or serial number
More than 100 numbers, contact manufacturer.
Model or catalog number
186-1046
Companies
- Manufacturer
-
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES