Health product recall

Brilliance CT Big Bore System-Main Unit, Brilliance ICT System - System, Brilliance ICT SP System - Main and Ingenuity CT System (2018-02-14)

Starting date:
February 14, 2018
Posting date:
March 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66128

Affected products

  1. BRILLIANCE CT BIG BORE SYSTEM-MAIN UNIT
  2. BRILLIANCE ICT SYSTEM- SYSTEM
  3. BRILLIANCE ICT SP SYSTEM - MAIN
  4. INGENUITY CT SYSTEM

Reason

Upon closing the Emergency Stop (E-STOP) the system performs an internal check. As a result of the motion controller in the couch, there is a delay in the response to the internal check. This delay may cause the couch carbon top to enter a free float state, which may allow it to move during patient load or unload activities. This free float motion results in the ESTOP being opened due to the unexpected couch motion. This issue occurs frequently after closing E-STOP. The motion controller in the couch detects a power problem and disables the horizontal motor, which allows the couch top to free-float*.                                         

*The term free float refers to the couch top being in a state that is not locked in position by the brake or being moved in a controlled fashion manual or automatic motion.

Affected products

A. BRILLIANCE CT BIG BORE SYSTEM-MAIN UNIT

Lot or serial number

Not applicable

Model or catalog number

4535 670 88051

728244

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 Miner Road

Cleveland

44143

UNITED STATES


B. BRILLIANCE ICT SYSTEM- SYSTEM

Lot or serial number

Not applicable

Model or catalog number

728306

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 Miner Road

Cleveland

44143

UNITED STATES


C. BRILLIANCE ICT SP SYSTEM - MAIN

Lot or serial number

Not applicable

Model or catalog number

728311

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 Miner Road

Cleveland

44143

UNITED STATES


D. INGENUITY CT SYSTEM

Lot or serial number

Not applicable

Model or catalog number

728323

728326

Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 Miner Road

Cleveland

44143

UNITED STATES