Health product recall

CellMek SPS

Last updated

Summary

Product
CellMek SPS
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
CellMek SPS All lots. C44603

Issue

Issue #1- The instructions for use states that the dead volume for tubes in cassette B is 750ul and cassette E is 1300ul. However, testing shows that the dead volumes are actually 1582ul and 2093ul.

Issue # 2- Cassette type E is not rocked prior to aspirating the specimen.

Recall start date: August 27, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-78025