Clip Gun System - Clip Gun (2017-12-18)
- Starting date:
- December 18, 2017
- Posting date:
- January 12, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65680
Affected products
Clip Gun System - Clip Gun
Reason
Testing performed by Medtronic showed that the outer sterile pouch of the product did not meet standards for package integrity. Specifically, the testing demonstrated that the outer pouch could be compromised and possibly be undetected by the user, which could potentially lead to patient infections. In an abundance of caution, these affected products are being removed from the field.
Affected products
Clip Gun System - Clip Gun
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- CG8900
- CG8901
Companies
- Manufacturer
-
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES