Health product recall

Color Cuff - Sterile Disposable Tourniquet Cuffs

Last updated

Summary

Product
Color Cuff - Sterile Disposable Tourniquet Cuffs
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Color Cuff - Sterile Disposable Tourniquet Cuffs More than 10 numbers, contact manufacturer. 5921-018-235
Color Cuff - Sterile Disposable Tourniquet Cuffs More than 10 numbers, contact manufacturer. 5921-018-135
Color Cuff - Sterile Disposable Tourniquet Cuffs More than 10 numbers, contact manufacturer. 5921-024-235

Issue

There is potential for the tourniquet cuffs flange to detach from the bladder where the welding connection is, which may result in the cuff not able to maintain air pressure. On February 12, 2025, a nonconformance was opened to address a manufacturing nonconformance identified through a Stryker complaint investigation for the tourniquet cuff flange becoming detached from the bladder. Less noticeable detached flanges have been reported and events confirmed.

Recall start date: June 27, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Stryker Instruments, A Division Of Stryker Corporation

1941 Stryker Way, Portage, Michigan, United States, 49002

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77715