Comprehensive® Shoulder System Primary Mini Length Shouldler Stem
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Comprehensive® Shoulder System Primary Mini Length Shouldler Stem |
65915809 |
113627 |
Issue
Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive® Shoulder Primary Stem (7mm Micro & 7 mm Mini). Five complaints reported that the stem implant did not fit in the inserter or that the inserter could not be detached from the stem. There were no injuries reported associated with the five complaints received. The internal investigation identified excess metal on the inner taper of the stem due to a chipped manufacturing tool.
Recall start date: March 10, 2025
Additional information
Details
Biomet Orthopedics
56 East Bell Drive, Warsaw, Indiana, United States, 46581
Get notified