Health product recall

Cook Medical Needles, Peel-Away® Introducers, and Dilators

Brand(s)
Last updated

Summary

Product
Cook Medical Needles, Peel-Away® Introducers, and Dilators
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected product

Lot or serial number

Model or catalogue number

Disposable Chib Biopsy Needles

More than 10 numbers, contact manufacturer.

DCHN-18-15.0
DCHN-22-10.0
DCHN-22-20.0
DCHN-18-10.0
DCHN-22-10.0-U
DCHN-22-15.0
DCHN-22-15.0-U

Peel-Away Introducer Sets And Trays

More than 10 numbers, contact manufacturer.

PLVW-XXX

Coons Dilator-Polyethylene

More than 10 numbers, contact manufacturer.

JCD12.0-38-20-COONS
JCD14.0-38-20-COONS
JCD10.0-38-20-COONS

Vessel Dilator-Polyethylene

More than 10 numbers, contact manufacturer.

JCD 8.0-38-20
JCD 8.0-38-15
JCD10.0-38-20
JCD 8.0-38-20-HC
JCD20.0-38-20
JCD22.0-38-20
JCD16.0-38-20-HC
JCD14.0-38-20-HC

Issue

Cook Medical identified that devices from the affected device lots may have packaging seals that do not meet peel strength specifications. This creates a potential of a sterile barrier failure that could result in compromised device sterility.

Recall start date: March 1, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - General and plastic surgery
Companies
Cook, Inc.
750 Daniels Way, Bloomington, Indiana, United States, 47404
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75601

Get notified

Receive notifications for new and updated recalls and alerts by category.