Health product recall

Digital Diagnostic Mobile X-Ray System GM85

Last updated

Summary

Product
Digital Diagnostic Mobile X-Ray System GM85
Issue
Medical devices - Performance
What to do

Customers to be contacted to arrange visit for inspection and repair. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Digital Diagnostic Mobile X-Ray System GM85 S/N 5143M3UJ200001N GM85

Issue

We have received a report regarding the loosening or breakage of bolts that connect the tube head unit (THU) to the arm. An investigation determined that this may be caused by metal fatigue resulting from accumulated stress over a long period of use. If all connecting bolts were to fail, it could potentially cause the THU to detach from the arm, posing a risk of serious injury to a patient or user.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Samsung Electronics Co., Ltd.

129, Samsung-Ro, Yeongtong-Gu, Suwon-Si, Gyeonggi-Do, South Korea, 16677

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78404

Get notified