Health product recall

directCHECK™ Whole Blood Control

Brand(s)
Last updated

Summary

Product
directCHECK™ Whole Blood Control
Issue
Medical devices - Labelling and packaging
What to do

Customer notification letter to be sent. Instructions to be provided on disposal of affected units

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
directCHECK™ Whole Blood Control A6DLA001 DCGLR‑2

Issue

Instances of consistently elevated results were identified when using directCHECK™ Whole Blood Control DCGLR‑2, lot A6DLA001. A process error led to testing the lot on the wrong instrument platform, which resulted in the creation and release of incorrect Quality Control (QC) ranges on the performance range card distributed with this lot.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies
Bionostics Inc
7 Jackson Road, Devens, Massachusetts, United States, 01434
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82075

Get notified

Receive emails about new and updated recall and safety alerts.