Health product recall

Elekta Modulator Discharge Rod

Last updated

Summary

Product
Elekta Modulator Discharge Rod
Issue
Medical devices - Performance
What to do

Elekta will issue an IFSM to check and replace faulty discharge rod crimps as needed. A service representative will contact users. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Elekta Infinity All lots. 1550358
Elekta Infinity All lots. 1552092
Versa HD All lots. 1550359
Versa HD All lots. 1552093
Harmony All lots. 1553032
Elekta Synergy All lots. 1550357
Elekta Synergy All lots. 1550291

Issue

Elekta has identified a safety concern related to the modulator discharge rod. It was discovered that the supplier had been incorrectly assembling the crimp within the crimping tool - fitting it in reverse. As a result, the earth lead on the safety discharge rod was able to be pulled out with minimal force, which could pose potential safety risks to the field service engineer. While not all crimps are affected, reversed crimps were found in stock at Elekta.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Elekta Solutions Ab
Hagaplan 4, Stockholm, Sweden, 11368
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81238