Extracorporeal Tubing Set for Heart-Lung Machine - Perfusion Pack and Art.Cannula - Wire Reinforced, Straight Tip W/Ring, Pre-Mounted Connector (2018-05-07)
- Starting date:
- May 7, 2018
- Posting date:
- June 1, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66946
Affected products
A. Extracorporeal Tubing Set for Heart-Lung Machine - Perfusion Pack
B. Art Cannula - Wire Reinforced, Straight Tip W/Ring, Pre-Mounted Connector
Reason
An arterial cannula used for open heart surgery is normally equipped with a vent plug, which facilitates the de-airing of the cannula. The vent plug of our arterial cannula consists of sintered polyethylene, therefore the structure of the vent plug is porous and gas-permeable. After the de-airing process is completed the customer has to separate the vent plug from the cannula to connect the extracorporeal circuit. In some rare cases it has come the vent plug broke during the separation from the arterial cannula.
Affected products
A. Extracorporeal Tubing Set for Heart-Lung Machine - Perfusion Pack
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
H 94500
Companies
- Manufacturer
-
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
B. Art Cannula - Wire Reinforced, Straight Tip W/Ring, Pre-Mounted Connector
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
A20-0106
A20-0107
A22-0106
A22-0107
A24-0107
Companies
- Manufacturer
-
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY