FilmArray Blood Culture Identification Panel (BCID) Panel (2018-09-06)
- Starting date:
- September 6, 2018
- Posting date:
- October 5, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67878
Affected products
FilmArray Blood Culture Identification Panel (BCID) Panel
Reason
BioFire Diagnostics, LLC has identified an increased risk of false positive proteus results when the Filmarray Blood Culture Identification (BCID) panel is used with BacT/ALERT Culture media with expiration dates of 2019-05-02 and beyond. The intended function of BacT/ALERT Blood culture bottles (culturing viable microorganisms) is not compromised by this anomaly. It is unknown if subsequent lots of media will also be subject to this risk.
Affected products
FilmArray Blood Culture Identification Panel (BCID) Panel
Lot or serial number
All lots.
Model or catalog number
RFIT-ASY-0126
RFIT-ASY-0127
Companies
- Manufacturer
-
BioFire Diagnostics, LLC
515 Colorow Dr
Salt Lake City
84108
Utah
UNITED STATES