Health product recall

FilmArray Blood Culture Identification Panel (BCID) Panel (2018-09-06)

Starting date:
September 6, 2018
Posting date:
October 5, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67878

Affected products

FilmArray Blood Culture Identification Panel (BCID) Panel

Reason

BioFire Diagnostics, LLC has identified an increased risk of false positive proteus results when the Filmarray Blood Culture Identification (BCID) panel is used with BacT/ALERT Culture media with expiration dates of 2019-05-02 and beyond. The intended function of BacT/ALERT Blood culture bottles (culturing viable microorganisms) is not compromised by this anomaly. It is unknown if subsequent lots of media will also be subject to this risk.

Affected products

FilmArray Blood Culture Identification Panel (BCID) Panel

Lot or serial number

All lots.

Model or catalog number

RFIT-ASY-0126

RFIT-ASY-0127

Companies
Manufacturer

BioFire Diagnostics, LLC

515 Colorow Dr

Salt Lake City

84108

Utah

UNITED STATES