FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
Brand(s)
Last updated
Summary
Product
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
FreeO2 Automated Oxygen Therapy Device Adult & Pediatric |
Version HW 1.2 |
FO2-110-00 |
Issue
A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
| Oxynov, Inc. |
| 725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82735
Get notified
Receive emails about new and updated recall and safety alerts.