hCG Cassettes
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
hCG Combo Cassette 25 |
All lots. |
92225CA |
hCG Cassette 25 |
All lots. |
92217CA |
Issue
Product may detect low levels of HCG below the stated assay sensitivity which may result in false positive test results in samples from non-pregnant individuals. The package insert is being updated to note that low HCG levels can occur at or after onset of perimenopause. Customers should consider this when interpreting results.
Additional information
Details
Abbott Rapid Diagnostics Ulc
185 Corkstown Road, Ottawa, Ontario, Canada, K2H 8V4