Histocore Pegasus
Brand(s)
Last updated
Summary
Product
Histocore Pegasus
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Histocore Pegasus |
More than 10 numbers, contact manufacturer. |
14048858009 |
Issue
Leica Biosystems became aware, through post market surveillance, of an issue regarding poorly processed or damaged biopsy tissue specimens on the Histocore Pegasus / Histocore Pegasus Plus, resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks. This recall is to inform customers of this issue and actions to be taken.
Recall start date: April 3, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Pathology
Companies
Leica Biosystems Nussloch Gmbh |
Heidelberger Street 17-19, Nussloch, Germany, 69226 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75407