Horizon® Microclip™ Titanium Ligating Clips with Tape
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Horizon® Microclip™ Titanium Ligating Clips with Tape |
More than 10 numbers, contact manufacturer. |
005200 |
Issue
Teleflex is initiating this voluntary recall because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures. Additionally, this introduces a further risk to the patient whereby the tips of the applier may project further past the tissue to be ligated, potentially risking injury to other distal anatomical structures.
Recall start date: May 2, 2023
Additional information
Details
Teleflex Medical
3015 Carrington Mill Blvd., Morrisville, North Carolina, United States, 27560
Get notified