Product
Impella Controller
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Impella Controller IC2034 0042-0000-CA
Impella Controller IC1572 0042-0000-CA
Impella Controller IC2369 0042-0000-CA
Impella Controller IC2370 0042-0000-CA
Impella Controller IC1715 0042-0000-CA
Impella Controller IC4944 0042-0000-CA
Impella Controller IC1480 0042-0000-CA
Impella Controller IC1544 0042-0000-CA
Impella Controller IC2032 0042-0000-CA
Impella Controller IC1665 0042-0000-CA
Impella Controller IC1098 0042-0000-US

Issue

Abiomed has decided to initiate a voluntary medical device recall (removal) for Automated Impella Controllers (AICs) in the field related to a design change that improves the AIC purge retainer fixture and addresses potential purge retainer failures. Failures impacting purge pressure and detection of the purge disc may lead to errors, alarms, and potential interruptions in hemodynamic support when the AIC may need to be replaced.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Abiomed Inc.
22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Recall date
Identification number
RA-78096

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe