Health product recall

Impella® Controller

Brand(s)
Last updated

Summary

Product
Impella® Controller
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Impella® Controller IC1544 0042-0000-CA
Impella® Controller IC1543 0042-0000-CA

Issue

Abiomed has identified specific automated Impella® controller in the field that have a pump driver circuit assembly that does not meet current specifications. These pump driver circuit assemblies contain 25v-rated tantalum capacitors instead of 35v-rated. These underrated capacitors may lead to decreased pump performance or pump stop and trigger an "Impella failure" or "Impella stopped. Controller failure." alarm.

Recall start date: August 27, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Abiomed Inc.

22 Cherry Hill Drive, Danvers, Massachusetts, United States, 01923

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-78001

Get notified

Receive notifications for new and updated recalls and alerts by category.