Inceptiv
Brand(s)
Last updated
Summary
Product
Inceptiv
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Inceptiv | More than 10 numbers, contact manufacturer. | 977119 |
Issue
Performing a switch device operation from the 97725 or 9772501 WENS using the A71300 SCS trialing app to transfer data (e.g. patient and lead information) to a new implanted Inceptiv or Intellis Pro Implantable Neurostimulator (INS) results in the A71400 CP app crashing and the INS becoming unusable when the transfer involves a predefined legacy lead option selected in the app that uses the forward slash character "/".
Recall start date: May 28, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Medtronic Inc. |
710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77580
Get notified