Insufflation Filter, Sterile (2021-05-21)
- Starting date:
- May 21, 2021
- Posting date:
- June 3, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-75723
Last updated:
2021-06-03
Affected Products
Insufflation Filter, Sterile
Reason
KARL STORZ was informed about potential deviations of validated parameters for ethylene oxidesterilization at sterilization provider Steril Milano. The deviations affect certain production lots of KARLSTORZ's Insufflation Filter 031122-25 and occurred between March 2018 and February 2021.
Affected products
Insufflation Filter, Sterile
Lot or serial number
18L0474FAX
Model or catalog number
031122-25
Companies
- Manufacturer
-
KARL STORZ SE & CO. KG
DR.-KARL-STORZ STRASSE 34
TUTTLINGEN
78532
GERMANY