MassCheck® and 3PLUS1® Antimycotic Drugs/Extended Plasma Controls
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
MassCheck® Antimycotic Drugs/Extended Plasma Control Level I |
4922 |
0253/XT |
3PLUS1® Multilevel Plasma Calibrator Set MassTox® Antimycotic Drugs/Extended |
4922 |
92051/XT |
MassCheck® Antimycotic Drugs/Extended Plasma Control Level II |
4922 |
0254/XT |
Issue
New data from stability tests can not confirm Chromsystems instruments & Chemicals GmbH previous information regarding the storage life of calibrator 92051/XT and controls 0253/XT, 0254/XT after reconstitution at storage temperatures of +20 to +25 °c and +2 to +8 °c provided in the instructions for use and on the package inserts. In addition, adsorption effects may occur if the reconstituted products are stored in glass containers.
Recall start date: February 5, 2025
Additional information
Details
Chromsystems Instruments & Chemicals GmbH
Am Haag 12, Grafelfing, Germany, 82166