Monaco (2020-04-09)
- Starting date:
- April 9, 2020
- Posting date:
- May 15, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73075
Last updated: 2020-05-15
Affected Products
Monaco
Reason
When editing a contour and clicking on a coronal or sagittal slice, the 3D volume is recreated and re-sliced when it should not be.The result can be a change in contour shape and volume. Therefore, the patient anatomy might not be accurately represented. The problem can happen with reshape contour and replace contour as well as any contouring tool that will activate a contour when the user clicks on the view.
Affected products
Monaco
Lot or serial number
- Version 5.40
- Version 5.40.01
Model or catalog number
- Monaco RTP System
Companies
- Manufacturer
-
Elekta INC.
13723 RIVERPORT DRIVE,, SUITE 100
Maryland Heights
63043
Missouri
UNITED STATES