Health product recall

Monnal™ T60 Advanced

Last updated

Summary

Product
Monnal™ T60 Advanced
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
Monnal™ T60 Advanced MT6A-02322 KA017124
Monnal™ T60 Advanced MT6A-02324 KA017124
Monnal™ T60 Advanced MT6A-02323 KA017124
Monnal™ T60 Advanced MT6A-02325 KA017124

Issue

Air Liquide Medical Systems has identified a manufacturing non-conformity that affects a limited and specific number of Monnal T60 and Monnal T60 advanced devices. It concerns the failure to fit washers beneath the screw head that secures the inspiratory unit of the device. The absence of a washer calls into question the conformity of the Monnal T60 units concerned with the requirements of standard EM 13718-1 (Requirements for medical devices used in air ambulances) and standard EN 1789+A1 (Requirements for medical devices used in road ambulances).

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Air Liquide Medical Systems

6 Rue Georges Besse, Antony, France, 92182

Published by
Health Canada
Audience
Industry
Recall class
Type II
Recall date
Identification number
RA-81994

Get notified

Receive emails about new and updated recall and safety alerts.