Health product recall

Monnal™ T60 Ventilator

Last updated

Summary

Product
Monnal™ T60 Ventilator
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
Monnal™ T60 Ventilator Not applicable. KA017115
Expiratory Valve Monnal EVA V2 Not applicable. KY732991

Issue

The Monnal EVA V2 (2nd generation - expiratory valve) used on Monnal T60/T60 Advanced ventilators can lead to under-delivery of volumes in "child/infant" patients due to poor membrane sealing. No patient incidents have been reported; the detection occurred during tests. Air liquide medical systems is recalling these valves to replace them with the first generation to prevent the risk of hypoventilation.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Air Liquide Medical Systems

6 Rue Georges Besse, Antony, France, 92182

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-78347