Health product recall

myAirvo™ 2 and Airvo™ 2 Humidifier

Last updated

Summary

Product
myAirvo™ 2 and Airvo™ 2 Humidifier
Issue
Medical devices - Performance
What to do

Customers will be notified with instructions to follow.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
myAirvo™ 2 Humidifier 09420012422248 PT100US
Airvo™ 2 Humidifier 09420012422347 PT101US

Issue

This recall is to inform customers of an update to the disinfection kit user manual for maintaining the Airvo™ and myAirvo™ 2 device throughout its expected service life. The Airvo™ 2 and myAirvo™ 2 device has an audible alert that sounds for at least 120 seconds when the unit is unintentionally disconnected from a power source. The disinfection kit user manual has been updated to include a regular test of this alert, to be carried out in between patient uses.  This update has been made due to reports of the audible alert sounding for less than 120 seconds, which may delay user awareness to reconnect power. There have been no events with serious injuries or death related to this issue.

Recall start date: August 27, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Fisher And Paykel Healthcare Ltd.

15 Maurice Paykel Place East Tamaki, Auckland, New Zealand, 2013

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-78053

Get notified

Receive notifications for new and updated recalls and alerts by category.