Health product recall

Neuro Omega system (2018-09-04)

Starting date:
September 4, 2018
Posting date:
October 5, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67880

Affected products

Neuro Omega system

Reason

In case Neuro Omega "Global stimulation return" connector of the EMG/ECOG Headbox is not properly connected, stimulation current may return through the Microelectrodes' contacts (Micro and Macro). Current can reach high charge density which may cause tissue harm.

Affected products

Neuro Omega system

Lot or serial number

Serial #: 173/06, 183/06,190

Model or catalog number

NEURO OMEGA

Companies
Manufacturer

Alpha Omega Engineering LTD

Nazareth Industrial Park Building, Mount Precipice, P.O. Box 2268

Nazareth

1612102

ISRAEL