Health product recall

Novasight Hybrid Catheter

Last updated

Summary

Product
Novasight Hybrid Catheter
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Novasight Hybrid Catheter

240502

TA-06-0001

Issue

A portion of the catheter was detached and left in the patient's body during a clinical case in USA on Nov.21, 2024. The portion was retrieved by the physicians, and there was no patient injury reported. 

Health Canada FRN was submitted on Mar.8, 2024 following initiation of a voluntary recall in USA. 

As there is a probability that use of or exposure to the affected catheter lots may cause adverse health consequences, Conavi Medical Inc. proposes a voluntary recall of Novasight Hybrid Catheters in Canadian market starting March 12, 2025.

Recall start date: Marc 12, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Conavi Medical Inc.
293 Lesmill Road, Toronto, Ontario, Canada, M3B 2V1
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77142

Get notified

Receive notifications for new and updated recalls and alerts by category.