Product
Omnilab Advanced, Domestic and Domestic Core
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

Omnilab Advanced, Domestic Core

All lots.

1111123
1111124

Omnilab Advanced, Domestic

All lots.

1111122

Issue

The affected products feature a ventilator inoperative alarm, which occurs when the ventilator detects an internal error or a condition that may affect therapy. this may manifest in the following ways:

• The device may reboot intermittently for 5-10 seconds, restarting therapy, and returning to delivering therapy with same patient settings.

• The device may reboot intermittently for 5-10 seconds, restarting therapy, and returning to delivering therapy but with factory default settings.

• When there are three (3) reboots within a 24-hour period, the device will enter a ventilator inoperative state.

• The device may enter a ventilator inoperative state without a reboot preceding this condition.

Any of the above scenarios could result in interruption and/or loss of therapy which may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.

Recall Start Date: March 25, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Respironics Inc. Also Trading As Philips Rs North America Llc

1001 Murry Ridge Lane, Murrysville, Pennsylvania, United States, 15668

Published by
Health Canada
Audience
General public
Recall class
Type II
Identification number
RA-75436

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe