Omnilab Advanced, Domestic and Domestic Core
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Omnilab Advanced, Domestic Core |
All lots. |
1111123 |
Omnilab Advanced, Domestic |
All lots. |
1111122 |
Issue
The affected products feature a ventilator inoperative alarm, which occurs when the ventilator detects an internal error or a condition that may affect therapy. this may manifest in the following ways:
• The device may reboot intermittently for 5-10 seconds, restarting therapy, and returning to delivering therapy with same patient settings.
• The device may reboot intermittently for 5-10 seconds, restarting therapy, and returning to delivering therapy but with factory default settings.
• When there are three (3) reboots within a 24-hour period, the device will enter a ventilator inoperative state.
• The device may enter a ventilator inoperative state without a reboot preceding this condition.
Any of the above scenarios could result in interruption and/or loss of therapy which may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.
Recall Start Date: March 25, 2024