Health product recall

Ortho Connect

Last updated

Summary

Product
Ortho Connect
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Ortho Connect

3.2.1
3.2.3
3.0.1
3.01
3.2.2
3.2.4
3.2

6904412
6904411

Issue

An assay rule was created for profile anti-A that was defined with assay anti-A has a grade between 0 and <= 3+ to hold pending review and to notify. When the result for anti-A was received with a grade of 3+, the result did not trigger the rule as was expected.  

Recall Start Date: April 1, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Biomedical Data Solutions Limited

Millenium Court, Burton On Trent, Staffordshire, United Kingdom, DE14 2WH

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77428

Get notified

Receive notifications for new and updated recalls and alerts by category.