PAT CBL 4WIRE AHA SNAP HC (2020-04-23)
- Starting date:
- April 23, 2020
- Posting date:
- May 15, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73097
Last updated: 2020-05-15
Affected Products
PAT CBL 4WIRE AHA SNAP HC
Reason
Internal testing has indicated that, in extremely rare cases, impacted Welch Allyn products may not meet the "Defibrillation Withstand" requirements of EN/IEC 60601-2-25 Medical Electrical Equipment. These are particular requirements for the Safety of Electrocardiographs; a standard the product claims to meet.
Affected products
PAT CBL 4WIRE AHA SNAP HC
Lot or serial number
- Not applicable.
Model or catalog number
- 9293-032-52
Companies
- Manufacturer
-
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES