Health product recall

Plum 360 Infusion System With Mednet

Last updated

Summary

Product
Plum 360 Infusion System With Mednet
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Plum 360 Infusion System With Mednet

More than 10 numbers, contact manufacturer.

30010

Issue

Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions. Alarms on Plum 360 consist of three elements: a visual alarm indication, an audible signal, and a message that appears on the display. Plum 360 pumps test the primary alarm audible signal during the power-on sequence. If the pump detects a failure of the audible signal during power-on, the pump display will remain off and the line b drip indicator will flash, indicating the pump has failed power-on. If the audible signal fails after passing the power-on test and during the use of the pump, a situation may occur where the pump has stopped infusion and an alarm is triggered without an audible signal. The pump will remain in this state until a clinician addresses the alarm. The visual alarm indication and the message will still appear on the display.

 

Recall start date: July 5, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Icu Medical, Inc.
600 North Field Drive, Lake Forest, Illinois, United States, 60045
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73967