Health product recall

RAPIDPoint 405, 500 and RAPIDLab 1245 Blood Gas Analyzer (2018-07-24)

Starting date:
July 24, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67392

Affected products

  1. RAPIDPoint 405 Blood Gas Analyzer
  2. RAPIDLab 1245 Blood Gas Analyzer
  3. RAPIDPoint 500 Blood Gas Analyzer

Reason

Siemens Healthcare Diagnostics has determined that 1 mg/ml and 2 mg/ml of hydroxocobalamin may interfere with Total hemoglobin (tHb) and some of the co-ox fractions that are reported on the RAPIDPoint 405, RAPIDPoint 500, RAPIDLab1245 and RAPIDLab1265 blood gas analyzers.

There have not been any reports of injury due to this issue. Siemens is issuing an Urgent Field Safety Notice (UFSN) to inform customers of this potential interfering substance.

Affected products

A. RAPIDPoint 405 Blood Gas Analyzer

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number
  • 05320281
  • 129887
  • RAPIDPOINT 405
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES


B. RAPIDLab 1245 Blood Gas Analyzer

Lot or serial number

10321844

10337179

10491395

Model or catalog number

1245

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES


C. RAPIDPoint 500 Blood Gas Analyzer

Lot or serial number

10492730

10696855

10696857

10697306

Model or catalog number

10492730

RAPIDPOINT 500

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES