RayStation & RayPlan (2019-11-18)
- Starting date:
- November 18, 2019
- Posting date:
- January 20, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72109
Last updated: 2020-01-21
Affected Products
- RayStation
- RayPlan
Reason
Editing a static arc beam may unexpectedly set the MU of the beam to its initial default value of 200 MU. This may happen in the following scenarios:
1. When editing or viewing a static arc beam with a cone using the edit beam dialog. The bug is triggered when closing the edit beam dialog with ok, even if no changes were made in the dialog.
2. When editing the aperture of a static arc beam using the polygon tool or the rectangular field tool.
If the bug is triggered, the dose is invalidated and needs to be recalculated before the plan can be approved or exported. After recalculation the dose will be correct, but the beam weighting can be different than it was before the beam was edited, even if the beam weights were clamped (locked) in the beam weighting tab.
Affected products
A. RayStation
Lot or serial number
All lots.
Model or catalog number
4.0.3
5
6
7
Companies
- Manufacturer
-
RaySearch Laboratories AB (PUBL)
Sveavagen 25
Stockholm
11134
SWEDEN
B. RayPlan
Lot or serial number
All lots.
Model or catalog number
2
7
Companies
- Manufacturer
-
RaySearch Laboratories AB (PUBL)
Sveavagen 25
Stockholm
11134
SWEDEN