RELIANCE defibrillation leads
Brand(s)
Last updated
Summary
Product
RELIANCE defibrillation leads
Issue
Medical devices - Performance
What to do
Healthcare professionals to conduct review of HVSI and determine if lead replacement is necessary.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
RELIANCE 4-FRONT | More than 10 numbers, contact manufacturer. | 0696 |
RELIANCE 4-FRONT | More than 10 numbers, contact manufacturer. | 0686 |
RELIANCE 4-FRONT | More than 10 numbers, contact manufacturer. | 0692 |
RELIANCE 4-FRONT | More than 10 numbers, contact manufacturer. | 0683 |
RELIANCE 4-FRONT | More than 10 numbers, contact manufacturer. | 0693 |
RELIANCE 4-FRONT | More than 10 numbers, contact manufacturer. | 0695 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0175 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0184 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0187 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0165 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0185 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0295 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0296 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0186 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0286 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0164 |
ENDOTAK RELIANCE G | More than 10 numbers, contact manufacturer. | 0167 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0181 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0292 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0161 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0282 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0283 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0170 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0293 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0171 |
ENDOTAK RELIANCE SG | More than 10 numbers, contact manufacturer. | 0180 |
Issue
Boston Scientific is initiating an advisory for ePTFE covered shock coil models from the RELIANCE family of defibrillation leads. The ePTFE material covering the shocking coils creates a matrix that, in some patients, promotes calcification, a biological wound response which is associated with gradual rising shock lead impedance (SLI) over time. Boston Scientific no longer manufactures or distributes ePTFE leads. Defibrillation leads currently distributed are non-ePTFE. Product is not being recalled.
Recall start date: July 24, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Boston Scientific Corporation |
4100 Hamline Avenue North, St. Paul, Minnesota, United States, 55112-5798 |
Published by
Health Canada
Audience
General public
Healthcare
Recall class
Type I
Identification number
RA-77782
Get notified
Receive notifications for new and updated recalls and alerts by category.