SERVO-i Ventilator System (2018-03-20)
- Starting date:
- March 20, 2018
- Posting date:
- April 13, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66400
Affected products
SERVO-i Ventilator System
Reason
Under certain conditions, an internal communication error might prevent the safety valve to close automatically. As a consequence, ventilation will stop a high priority alarm, as well as clinical alarms, will be triggered. We have no reports of any permanent injuries. This issue will stop ventilation, the safety valve open, thus allowing the patient to breathe freely, but without ventilator support. The high priority technical error alarm TE 11 will be activated simultaneously with other clinical alarms.
Affected products
SERVO-i Ventilator System
Lot or serial number
83775 - 87140
Model or catalog number
64 87 800 E407E
Companies
- Manufacturer
-
Maquet Critical Care AB
Rontgenvagen 2
Solna
17154
SWEDEN