Siemens Healthineers AG Magnetom and Biograph One Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Magnetom Sola |
More than 10 numbers, contact manufacturer. |
11291455 |
Magnetom Sola Fit |
More than 10 numbers, contact manufacturer. |
11410482 |
Magnetom Flow.Elite |
234786 |
11851911 |
Magnetom Flow.Elite |
234723 |
11851911 |
Magnetom Flow.Elite |
234827 |
11851911 |
Magnetom Flow.Elite |
234832 |
11851911 |
Magnetom Lumina |
196680 |
11344916 |
Biograph One |
238015 |
11689172 |
Biograph One |
238023 |
11689172 |
Magnetom Vida Fit |
202203 |
11410481 |
Magnetom Vida Fit |
202080 |
11410481 |
Magnetom Vida |
More than 10 numbers, contact manufacturer. |
11060815 |
Magnetom Viato.Mobile |
239040 |
11689652 |
Issue
A software issue has been identified in syngo.via MR Cardiac Analysis MR Flow Quantification affecting the evaluation of retrospectively gated 2D cardiac flow data.
- The algorithm does not correctly represent the final phase of the cardiac cycle. The evaluation stops at the last trigger time point (TT point), which marks the beginning of the final cardiac phase, instead of accounting for the complete duration of that phase until the end of the cardiac cycle. The final repetition-time interval (TR) is not included in the calculation (“TR missing”). This results in systematic truncation of the cardiac cycle and incomplete loop closure.
- Consequently, flow quantification parameters (e.g., Forward volume, Avg.Flow/Min) are systematically underestimated.
- The impact is relevant for Beat Sensor-triggered flow data, for which the flow peak occurs in the very last acquired cardiac phases.