Smoothbore Breathing System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
22mm Smoothbore Adult, Cough Assist Breathing System With Mask And Flo-Guard System |
32510470 |
5001011 |
Smoothbore Breathing System |
32527207 |
5183000 |
Smoothbore Breathing System |
32517823 |
5183000 |
Smoothbore Breathing System |
32516907 |
5183000 |
Smoothbore Breathing System |
32511321 |
5183000 |
Smoothbore Breathing System |
32527207 |
5191000 |
Smoothbore Breathing System |
32517823 |
5191000 |
Smoothbore Breathing System |
32516907 |
5191000 |
Smoothbore Breathing System |
32511321 |
5191000 |
Bilevel Breathing Systems |
32517459 |
5805044 |
Bilevel Breathing Systems |
32518657 |
5805044 |
Bilevel Breathing Systems |
32515237 |
5805044 |
CPAP Breathing System |
32511639 |
5804000 |
Issue
During manufacturing, insufficient tetrahydrofuran (THF) solvent is used to bond the soft connectors onto the Smoothbore tube. Following reports received from the market of disconnection of the soft connectors from the tube, our investigation and testing of internal stock has identified insufficient amount of solvent used to ensure a secure connection.