Health product recall

SPECTRALIS® with Flex Module

Last updated

Summary

Product
SPECTRALIS® with Flex Module
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

SPECTRALIS® with Flex Module

Spec-21986

S2610

SPECTRALIS® with Flex Module

Spec-24546

S3610

SPECTRALIS® with Flex Module

Spec-23776

S3610

SPECTRALIS® with Flex Module

Spec-13403

S3610

SPECTRALIS® with Flex Module

Spec-15400

S3610

Issue

There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose. As a result, the Flex stand may tip over. Users shall schedule a service appointment as soon as possible so that the castor fastenings can be inspected and corrected.
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Ophthalmology
Companies

Heidelberg Engineering GmbH

Max-Jarecki-Strasse 8, Heidelberg, Germany, 69115

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82505

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe