StealthStation S7 Cranial Software (2018-10-03)

Starting date:
October 3, 2018
Posting date:
November 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68088

Affected products

StealthStation S7 Cranial Software

Reason

The purpose of this notification is to provide information related to potential inaccuracy during biopsy procedures using the StealthStation S7 depth gauge feature in navigated biopsy procedures, where the "navigate projection" feature is used and the projection is longer than the length of the plan, the graphical biopsy depth gauge feature can display inaccurate information. The biopsy needle depth gauge is a numerical and graphical representation of the needle cutting window in the guidance view quadrant to assist in visualization of the cutting window position.

During a biopsy procedure, the monitor shows four quadrants (trajectory 1, trajectory 2, guidance, and probe's eye views) as well as the biopsy depth gauge. The overlay of the biopsy needle within the anatomical views accurately indicates the correct tip location of the instrument. However, if "navigate projection" is selected, and the projection is longer than the length of the plan, then the biopsy depth gauge graphical display inaccurately depicts that the tip of the instrument has not yet reached the plan target.

Affected products

StealthStation S7 Cranial Software

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 9733763
Companies
Manufacturer

Medtronic Navigation Inc.