Health product recall

SYNCHRON Systems Lipase (LIP) Reagent (2018-09-11)

Starting date:
September 11, 2018
Posting date:
October 5, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67886

Affected products

SYNCHRON Systems Lipase  (LIP) Reagent

Reason

Control and patient samples tested with lip reagent lot M712448 may recover lower than those tested with other lots. Samples below 100 U/L may recover up to 10.7 U/L lower. Samples above 100 U/L may recover up to 8.9% lower. Lower recovery of lip in control samples may exceed the lower limits of a laboratory's established QC range or published insert range.

Affected products

SYNCHRON Systems Lipase  (LIP) Reagent

Lot or serial number

M712448

Model or catalog number

476851

Companies
Manufacturer

Beckman Coulter Inc.