SYNCHRON Systems Lipase (LIP) Reagent (2018-09-11)
- Starting date:
- September 11, 2018
- Posting date:
- October 5, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67886
Affected products
SYNCHRON Systems Lipase (LIP) Reagent
Reason
Control and patient samples tested with lip reagent lot M712448 may recover lower than those tested with other lots. Samples below 100 U/L may recover up to 10.7 U/L lower. Samples above 100 U/L may recover up to 8.9% lower. Lower recovery of lip in control samples may exceed the lower limits of a laboratory's established QC range or published insert range.
Affected products
SYNCHRON Systems Lipase (LIP) Reagent
Lot or serial number
M712448
Model or catalog number
476851
Companies
- Manufacturer
-
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES