T2 Proximal Humeral Nail System - Long Nail (2018-06-15)
- Starting date:
- June 15, 2018
- Posting date:
- August 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67376
Affected products
T2 Proximal Humeral Nail System - Long Nail
Reason
The manufacturer has determined that products have left the factory where the quality of the sealed seam of the sterile packaging did not correspond to the specification. During the sealing process, a tyvek lid stuck to the sealing plate. The subsequently packaged products were sealed with the lid sticking to the seal plate.
Though testing of the sealed seam confirmed that the integrity of the sterile barrier is intact, integrity of the sterile barrier cannot be assured over 5 years shelf life.
Affected products
T2 Proximal Humeral Nail System - Long Nail
Lot or serial number
K0ACFED
K0B9038
K0B906C
K0BF6C6
K0BF6D2
K0CAEC4
K0CC741
K0CC742
K0CC743
K0CC744
Model or catalog number
1832-3826S
Companies
- Manufacturer
-
Stryker Trauma GMBH
Prof.-Kuntscher-Str.1-5
Schoenkirchen
24232
GERMANY