Health product recall

T2 Proximal Humeral Nail System - Long Nail (2018-06-15)

Starting date:
June 15, 2018
Posting date:
August 17, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67376

Affected products

T2 Proximal Humeral Nail System - Long Nail

Reason

The manufacturer has determined that products have left the factory where the quality of the sealed seam of the sterile packaging did not correspond to the specification. During the sealing process, a tyvek lid stuck to the sealing plate. The subsequently packaged products were sealed with the lid sticking to the seal plate.

Though testing of the sealed seam confirmed that the integrity of the sterile barrier is intact, integrity of the sterile barrier cannot be assured over 5 years shelf life.

Affected products

T2 Proximal Humeral Nail System - Long Nail

Lot or serial number

K0ACFED

K0B9038

K0B906C

K0BF6C6

K0BF6D2

K0CAEC4

K0CC741

K0CC742

K0CC743

K0CC744

Model or catalog number

1832-3826S

Companies
Manufacturer

Stryker Trauma GMBH

Prof.-Kuntscher-Str.1-5

Schoenkirchen

24232

GERMANY