Health product recall

t:slim X Insulin Pumps

Last updated

Summary

Product
t:slim X Insulin Pumps
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
t:slim X2 / Basal-IQ (Mmol/L) More than 10 numbers, contact manufacturer. 1006535
t:slim X2 / Basal-IQ (Mmol/L) More than 10 numbers, contact manufacturer. 1006419
t:slim/Control-IQ More than 10 numbers, contact manufacturer. 1005611
t:slim/Control-IQ, Replacement More than 10 numbers, contact manufacturer. 1005613

Issue

Tandem Diabetes Care, Inc. ("Tandem") has identified certain speaker versions that are in t:slim X insulin pumps have a higher rate of malfunction 16 events due to a wiring issue within the speaker.

Recall start date: August 18, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Tandem Diabetes Care, Inc.
12400 High Bluff Dr., San Diego, California, United States, 92130
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-77883

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe