Vanta/Sequentia Lt Clinician Programmer Application
Brand(s)
Last updated
Summary
Product
Vanta/Sequentia Lt Clinician Programmer Application
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Vanta/Sequentia Lt Clinician Programmer Application | More than 10 numbers, contact manufacturer. | A71200 |
Issue
The Vanta A71200 CP app V2.0.2465 and V2.0.2683 will display a "system error" if the user selects the start usage button twice when interrogating a new implantable neurostimulator (INS). This error prompts the user to restart the CP app and contact Medtronic for assistance. Upon subsequent interrogation of the INS, the Vanta A71200 CP app will display a "validation error" prompting the user to resolve the error in order to proceed with the INS interrogation. If the validation error is not resolved, the user is unable to complete the first attempt at interrogating the INS.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
| Medtronic Inc. |
| 710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81342