Health product recall

VENTANA® anti-Cd10 (SP67) Rabbit Monoclonal Primary Antibody

Last updated

Summary

Product
VENTANA® anti-Cd10 (SP67) Rabbit Monoclonal Primary Antibody
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

VENTANA® anti-Cd10 (SP67) Rabbit Monoclonal Primary Antibody

K14266
J17541
K09880
M00669
K06239
K26461
J11853
J25047
J04613
K00982
J302286
K19784

790-4506

Issue

Roche has received complaints regarding high background and off-target staining with certain lots of the VENTANA® anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody resulting in a risk of false positive results.

Subsequent investigation identified that this high background and off-target staining can interfere with slide interpretation when using the recommended method sheet OptiView DAB IHC Detection Kit (product no.760-700 / 06396500001) / OptiView Amplification Kit (product no. 760-099 / 06396518001) and the ultraView Universal DAB Detection Kit (product no.760-500 i 05269806001) /Amplification Kit (product no. 760-080 / 05266114001) protocol.

Recall start date: July 11, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Microbiology
Companies

Ventana Medical Systems Inc

1910 E Innovation Park Drive, Tucson, Arizona, United States, 85755

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75867

Get notified

Receive notifications for new and updated recalls and alerts by category.