Health product recall

VersaCross™ Access Solution, TorFlex™ Transseptal Guiding Sheath and VersaCross™ Transseptal Sheath

Last updated

Summary

Product
VersaCross™ Access Solution, TorFlex™ Transseptal Guiding Sheath and VersaCross™ Transseptal Sheath
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0031
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0122
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0137
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0034
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0123
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0132
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0133
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0102
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0131
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0022
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0024
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0121
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0125
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0134
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0021
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0025
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0032
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0124
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0101
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0103
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0111
VersaCross™ Access Solution More than 10 numbers, contact manufacturer. VXSK0112
TorFlex™ Transseptal Guiding Sheath 35644132 TF85-32-63-45
TorFlex™ Transseptal Guiding Sheath 35508747 TF85-32-63-45
TorFlex™ Transseptal Guiding Sheath 35746496 TF85-32-63-45
TorFlex™ Transseptal Guiding Sheath 35913449 TF85-32-63-45
TorFlex™ Transseptal Guiding Sheath 35738145 TF85-32-63-55
TorFlex™ Transseptal Guiding Sheath 35508748 TF85-32-63-55
VersaCross™ Transseptal Sheath 35800622 VXS0201

Issue

Boston Scientific is initiating a removal of specific lots of the VersaCross™ Access Solution, TorFlex™ Transseptal Guiding Sheath, and VersaCross™ Transseptal Sheath due to the potential for a hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility.

Recall start date: June 10, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Baylis Medical Company Inc.

5959 Trans-Canada Highway, Montreal, Quebec, Canada, H4T 1A1

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77630

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe