Health product recall

VITROS Chemistry Product System-Ca Slides (Calcium)

Last updated

Summary

Product
VITROS Chemistry Product System-Ca Slides (Calcium)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
VITROS Chemistry Product System-Ca Slides (Calcium) Gen 67 and above 145 0261

Issue

Generations (Gens) 67 and above of VITROS CA slides may experience an increased occurrence of condition code th4-63j during calibration, on VITROS XT 3400 and VITROS XT 7600 systems. Condition code th4-63j is generated when the reflectometer detects image variations which exceed an assay parameter defined on the assay data disk (add). If condition code th4-63j is generated during calibration, the calibration may fail.

Recall start date: July 14, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive, Rochester, New York, United States, 14626
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77764

Get notified

Receive notifications for new and updated recalls and alerts by category.