Health product recall

VITROS Chemistry Products PHYT Slides (2018-05-10)

Starting date:
May 10, 2018
Posting date:
June 6, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66970

Affected products

VITROS Chemistry Products PHYT Slides

Reason

Ortho confirmed customer complaints regarding the potential for imprecise results when using specific lots of the VITROS PHYT Slides.

Ortho performed internal testing to confirm the issue using several lots of coating 0165. The results obtained were either positively or negatively biased across the VITROS PHYT Slides measuring range of 3.00-40.00 ug/ml (11.88-158.40 umol/l).

Affected products

VITROS Chemistry Products PHYT Slides

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

8298671

Companies
Manufacturer

Ortho-Clinical Diagnostics Inc.

100 Indigo Creek Drive

Rochester

14626

New York

UNITED STATES