Health product recall

Vitros Chemistry Products System - CRP Slides (C-Reactive Protein)

Last updated

Summary

Product
Vitros Chemistry Products System - CRP Slides (C-Reactive Protein)
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalog number

Vitros Chemistry Products System - CRP Slides (C-Reactive Protein)

8812910052
8812910079
8812910019
8812910080

809 7990

Issue

During our internal testing, QuidelOrtho determined that coating 1291 was incorrectly assigned GEN 88, instead of the intended GEN 83. Product labeling for all lots manufactured from coating 1291 lists the generation as GEN 88, including the product carton, foil wrapper, and cartridge (cart) label. The cart label is scanned by the Vitros® System to determine the GEN of the product. The GEN is associated with specific calibration parameters, which are stored on assay data disks and calibration diskettes and used by Vitros systems to process Vitros slides.

Recall Start Date: April 25, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Immunology
Companies

Ortho-Clinical Diagnostics Inc.

100 Indigo Creek Drive, Rochester, New York, United States, 14626

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-77427