Health product recall

Vitros® Versatips and Vitros® Microtips

Last updated

Summary

Product
Vitros® Versatips and Vitros® Microtips
Issue
Medical devices - Performance issue
What to do

Review & post notice, discard affected assembly once replacement received, complete/return reply form. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Vitros 3600 Immunodiagnostics System (New & Refurbished)

More than 10 numbers, contact manufacturer.

680 2914
680 2783

Vitros XT 7600 Integrated System

More than 10 numbers, contact manufacturer.

684 4461

Vitros 4600 Chemistry System (New & Refurbished)

More than 10 numbers, contact manufacturer.

6900440
6802445

Vitros 4600 Chemistry System (New & Refurbished)

More than 10 numbers, contact manufacturer.

6900440
6802445

Vitros 5.1 FS Chemistry System (New & Refurbished)

More than 10 numbers, contact manufacturer.

6801375
6801890

Vitros XT 3400 Chemistry System

More than 10 numbers, contact manufacturer.

684 4458

Vitros 5.1 FS Chemistry System (New & Refurbished)

More than 10 numbers, contact manufacturer.

6801375
6801890

Vitros 5600 Integrated System (New & Refurbished)

More than 10 numbers, contact manufacturer.

680 2413
680 2915

Issue

Certain proboscis/piston assembly (or assemblies) manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an out-of-specification barrel component. This may cause Vitros® Versatips and/or Vitros® microtips to be improperly seated on the barrel component of the assembly, which may result in an increased frequency of condition codes related to the actions of tip pickup, aspirate, dispense, mix, tip-seal, and discard. due to the small Vitros® Microtip and Versatip® dimension variability (within manufacturing tolerances) and the way these tips interact with the assembly, Assemblies manufactured within the affected date range may not perform as intended based on the tips currently in use.

Recall start date: August 10, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Ortho-Clinical Diagnostics Inc.

100 Indigo Creek Drive, Rochester, New York, United States, 14626

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74209

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe